Stent grafts are composed of a metal stent and a polymer coating. They are used to reconstruct diseased vascular pathways. The coating is typically woven from materials such as PET, PE, and (e)PTFE.
The main types of woven coatings are woven and knitted.
Woven membranes offer a dense and uniform weave and ultra-thin thickness, demonstrating excellent strength, blood blockage, and release properties. They are widely used in medical devices for aorta, peripheral, and structural heart applications.
The interlocking coil structure of knitted membranes gives the fabric a high thickness, high elasticity, and excellent extensibility. They are suitable for filling, sealing, and fixation applications, and are widely used in valves and artificial blood vessels.
Based on these two core processes, we utilize our state-of-the-art equipment to develop a variety of sample sizes and provide proven membrane material solutions, helping customers shorten their R&D cycles and reduce development costs.
Applications
This membrane material can be widely used in medical devices, including:
Thoracic Aortic Stents | Abdominal Aortic Stents | Gastrointestinal Stents | Biliary Stents
Key Product Features
The fabric features a compact structure, high strength, small pore, low leakage, excellent deformation resistance, and excellent biocompatibility. Customized designs can also be provided to meet specific performance requirements.
Product Specifications
| Specifications | Water Permeability | Thickness | Tensile Strength (Axial) | Tensile Strength (Circumferential) | Burst Strength (Probe) |
|---|---|---|---|---|---|
| Customization | ≤ 200mL/(cm2·min) | 0.060mm~0.110mm | ≥6N/mm | ≥5.5N/mm | ≥100N |
| Specification | Water Permeability | Thickness | Burst Strength (Probe) | Suture Retention Strength |
|---|---|---|---|---|
20D45 | ≤ 300mL/(cm2·min) | 0.035mm~0.045mm | ≥ 150N | ≥ 1.0N |
40D55 | ≤ 200mL/(cm2·min) | 0.045mm~0.065mm | ≥ 150N | ≥ 1.0N |
40D70 | ≤ 300mL/(cm2·min) | 0.060mm~0.075mm | ≥ 150N | ≥ 1.0N |
40D85 | ≤ 300mL/(cm2·min) | 0.065mm~0.085mm | ≥ 150N | ≥ 1.0N |
100D60 | ≤ 300mL/(cm2·min) | 0.115mm~0.155mm | ≥ 150N | ≥ 1.0N |
42D55 | ≤ 500mL/(cm2·min) | 0.045mm~0.055mm | ≥ 150N | ≥ 1.0N |
42D55(P) | 实测 | 实测 | ≥ 150N | ≥ 1.0N |
30D70 | ≤ 200mL/(cm2·min) | 0.055mm~0.075mm | ≥ 150N | ≥ 1.0N |
30D70 (X) | ≤ 200mL/(cm2·min) | 0.045mm~0.070mm | ≥ 150N | ≥ 1.0N |
| Specification | Thickness | Tensile Strength (Axial) | Tensile Strength (Circumferential) | Burst Strength (Probe) |
|---|---|---|---|---|
| Customization | 0.10~0.40mm | ≥100N | ≥1.5N/mm | ≥50N |
| Specification | Thickness | Burst Strength (Probe) |
|---|---|---|
66A2 | 0.070~0.110mm | ≥ 50N |
66B2 | 0.080~0.120mm | ≥ 50N |
21B10 | 0.320~0.350mm | ≥ 50N |
30C4 | ≥0.200mm | ≥ 50N |
23B4 | 0.140~0.200mm | ≥ 50N |
55A3 | 0.070~0.120mm | ≥ 50N |
45B3 | 0.170~0.220mm | ≥ 50N |
Quality Assurance
Certified ISO 13485 Quality Management System
Class 10,000 (ISO Class 7) Cleanroom